Parameterization
104
Temperature adjustment
(ST / R / RS models)
This submenu facilitates to adjust the tolerances of the temperature probe in the measuring
cell (zero shift by 0.5 °C maximum).
Notes:
l
Because of the heat capacity of the measuring cell it is neccessary to place it into a water
vessel with at least 2 liters of water.
l
Leave measuring cell and reference thermometer at least 15 minutes, with temperature
differences between water and cell > 10°C at least 1 hour in this vessel and stir from time to
time until the adjustment can be made.
Setting:
With
select submenu
"PARA T-ADJ".
With
enter submenu
"PARA T-ADJ".
The adjustment is made with
in a
range of
∆∆ϑϑ = ±
0.5°C.
⇐
Field 2: Indication of the adjusted
temperature.
⇐
Field 3: Shows size of the adjustment.
The temperature in field 2 changes with the
adjustment input.
PARA T-REF
Submenu "PARA T-REF".
With
selection of further
submenus.
With
enter submenu
"PARA T-REF".
With
select between reference
temperature 20°C or 25°C.
With
go on to submenu "PARA T-C".