58
9 Analytical Quality Assurance (AQA)
Example: Report output
AQA check of meter
AQA1
26.08.99
10:23
Operator:
AQA interval
12 W.
AQA check of meter
ok
L check
ok
test solution
Measured value Unit
theor. value
tolerance
result
445-1
0.209
A
0.200
0.020
ok
445-2
0488
A
0.500
0.030
ok
445-3
0.986
A
1.000
0.040
ok
445-4
1.487
A
1.500
0.050
ok
525-1
0.210
A
0.200
0.020
ok
...