15 Updating method data
The following display appears:
Possible error messages
If a version of the method is already stored in the
photometer, a message appears together with the
dates of both versions:
→
Press
to stop the data transfer. The 15 minu-
tes warm-up period of the photometer begins
then.
→
Press
to select the transfer data menu item
→
Confirm with
.
The data of the MemoChip is transferred (continuation
display).
After the data transfer is complete, the following dis-
play appears:
The 15 minutes warm-up period of the photometer fol-
lows; the photometer then automatically performs a
SelfCheck.
Message
Meaning
Remedy
MemoChip unknown
MemoChip not designated for photo-
meter
→
Place the correct MemoChip in
the recess
MemoChip illegible
MemoChip defective or not program-
med
→
Request a new MemoChip
Wrong MemoChip
(sequence)
Sequence of the MemoChips incorrect
→
Ensure the correct sequence
Transmission error
Error in data transfer
→
Repeat the procedure