57
9 Analytical Quality Assurance (AQA)
After the measurement is finished:
Absorbance test OK...
...or error message
After approx. 3 s:
After approx. 5 s:
The last measurement is repeated.
→
Insert the next test solution
→
Cancel:
To cancel the check means no release for the next
“meter" AQA interval!
→
Continue until the photometer has measured all 12
test solutions.
Clearing the error:
→
Compare the theoretical absorbance to the lot val-
ues (see section 9.2.1
)
→
Repeat the measurement
(plug the cell in again)
→
Exchange the test solution (each packet contains
two identical test solutions)
→
If necessary, perform a zero adjustment and repeat
the check
→
Use a new Spectroquant
®
PhotoCheck packet
→
Quit and have the photometer checked in the factory
If the meter monitoring is canceled, the pho-
tometer is not released. Each time the pho-
tometer is switched on, the warning message
"Check meter" appears until the meter moni-
toring was successfully performed or the AQA
mode is switched off. After confirming with
the measuring can be continued as usu-
al.
Содержание PhotoLab Spektral
Страница 2: ......
Страница 31: ...Edition date 06 00 Order no ba 41111e04 Functional description Edition date 04 2004 Order no ba41111e08...
Страница 32: ......
Страница 38: ...8 List of contents...
Страница 128: ...98 13 User defined methods...
Страница 132: ...102 14 Photometer setup...
Страница 140: ...110 16 RS 232 C interface...
Страница 144: ...114 18 Exchanging the Memory Card...
Страница 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Страница 153: ...123 21 Appendix...
Страница 154: ...124 21 Appendix...
Страница 160: ......
Страница 161: ......