3 Commissioning
→
To switch off the photometer, close the cover.
After 5 seconds:
Automatic change to the concentration measuring
mode
The AutoCheck function automatically
checks and calibrates the optical measur-
ing unit. To do this, the photometer per-
forms cyclical measurements without a cell.
The photometer checks whether the signal
level of the optical unit in the meter lies with-
in the allowable tolerances. Furthermore,
the photometric zero point of the meter is
calibrated.
After the warm-up time, the first AutoCheck
is performed as part of the SelfCheck. From
this point in time, the AutoCheck is repeat-
ed every 5 minutes as long as no cell is in-
serted in the cell shaft. If an AutoCheck
cannot be performed because a cell is in-
serted, it is automatically repeated as soon
as the cell is removed.
During the measurement of the kinetic (see
section
), the photometer
cannot perform an AutoCheck due to the in-
serted cell. Thus, you must wait for a warm-
up period of 2 hours in this case.
After this time, the stability of the signals is
so high that the measuring accuracy is re-
tained over longer periods of time even
without an AutoCheck.
Содержание PhotoLab Spektral
Страница 2: ......
Страница 31: ...Edition date 06 00 Order no ba 41111e04 Functional description Edition date 04 2004 Order no ba41111e08...
Страница 32: ......
Страница 38: ...8 List of contents...
Страница 128: ...98 13 User defined methods...
Страница 132: ...102 14 Photometer setup...
Страница 140: ...110 16 RS 232 C interface...
Страница 144: ...114 18 Exchanging the Memory Card...
Страница 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Страница 153: ...123 21 Appendix...
Страница 154: ...124 21 Appendix...
Страница 160: ......
Страница 161: ......