71
10 Special functions
This function is used to newly determine the reference
absorbance or transmission for a wavelength range.
The baseline is stored and is used as a new reference
value for the spectrum measurement.
Erasing the baseline
The baseline can be erased by
●
using the menu item, erase baseline
●
switching off the meter.
→
Select the measure baseline function from the
spectrum menu.
→
Insert the cell to start the measurement.
The message measuring... appears on the display.
After the measurement is finished:
The baseline has been measured and stored.
→
Return to the spectrum menu using
.
→
Select the measure spectrum function from the
spectrum menu.
→
Measure as described in section
10.1.2
. While the
message measuring... is displayed, Baseline: OK
appears in the upper display line to indicate that
the measurement is being conducted against a
new baseline.
The measured baseline remains valid for all
the following spectrum measurements in the
same wavelength range and with the same
cell type.
10.1.3 Measuring the baseline
Содержание PhotoLab Spektral
Страница 2: ......
Страница 31: ...Edition date 06 00 Order no ba 41111e04 Functional description Edition date 04 2004 Order no ba41111e08...
Страница 32: ......
Страница 38: ...8 List of contents...
Страница 128: ...98 13 User defined methods...
Страница 132: ...102 14 Photometer setup...
Страница 140: ...110 16 RS 232 C interface...
Страница 144: ...114 18 Exchanging the Memory Card...
Страница 152: ...122 21 Appendix 21 1 Menu structure of the setup menu...
Страница 153: ...123 21 Appendix...
Страница 154: ...124 21 Appendix...
Страница 160: ......
Страница 161: ......