pH 3310
Operation
41
ba75802e02
12/2008
Note
For
single-point calibration
, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.
Continuing with two-
point calibration
(
Buffer
:
TEC)
6
Wait for the end of the measurement with stability control or
accept the calibration value with
<ENTER>
.
The calibration display for the next buffer appears.
7
If necessary, terminate the calibration as a single-point
calibration with
<M>
.
The calibration record is displayed.
pH
181.3
mV
24.8 °C
AutoCal
4.010
22.09.2008
08:00
AR
TEC
pH
2
181.3
mV
24.8 °C
AutoCal
22.09.2008
08:00
TEC
8
Thoroughly rinse the combination electrode with distilled water.
9
Immerse the electrode in buffer solution 2.
10
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with
<
▲
><
▼
>
.
11
Start the measurement with
<ENTER>
.
The measured value is checked for stability (stability control).