pH 3310
Operation
29
ba75802e02
12/2008
Criteria for a stable
measured value
The
Stability control
function checks whether the measured values are
stable within the monitored time interval.
The minimum duration until a measured value is assessed as stable is
the monitored time interval. The actual duration is mostly longer.
3
Release the frozen measured value again with
<AR>
or
<M>
.
The [AR] status display disappears. The display switches back
to the previous representation.
Measured
parameter
Time interval
Stability in the time interval
pH value
15 seconds
Δ
pH: better than 0.01