Operation
pH 3310
26
ba75802e02
12/2008
4.3.3
Automatic
Stability control
The automatic
Stability control
function continuously checks the
stability of the measurement signal. The stability has a considerable
impact on the reproducibility of measured values.
You can activate or switch off the automatic
Stability control
function
(see section 4.3.1).
The measured parameter flashes on the display
z
as soon as the measured value is outside the allowed stability range
z
if you switch over between the measured parameters with
<M>
.
z
if the automatic
Stability control
function is switched off.