pH 3310
Operation
47
ba75802e02
12/2008
4.4.9
Displaying calibration records
The calibration data can be displayed and then output to the interface.
Displaying the
calibration record
The calibration record of the last calibration is to be found under the
menu item,
Calibration
/
Calibration record
. To open it in the measured
value display, press the
<CAL
_
>
key.
The calibration records of the last calibration are to be found in the
menu,
Storage & config
/
Data storage
/
Calibration data storage
. To
open the
Storage & config
menu, press the
<F1
_
>
/[Menu] key in the
measured value display.
Menu item
Setting/
function
Description
Data storage
/
Calibration data
storage
/
Display
-
Displays the calibration
record.
Further options:
z
Scroll through the
calibration records with
<
▲
><
▼
>
.
z
Output the displayed
calibration record to the
interface with
<F2>
/
[USB output]
.
z
Quit the display with
<F1>
/
[Back]
or
<ENTER>
.
z
Switch directly to the
measured value display
with
<M>
.
Data storage
/
Calibration data
storage
/
Output to
USB
-
Outputs the calibration
records to the interface.