pH 3310
Operation
53
ba75802e02
12/2008
Starting the automatic
storage function
To start the automatic storage function, select
Continue
with
<
▲
><
▼
>
and confirm with
<ENTER>
. The meter switches to the measured value
display.
The active automatic storage function can be recognized from the
progress bar in the status line. The progress bar shows the remaining
storage duration.
Note
If the automatic storage function is activated, only the following keys
are active: Softkeys,
<M>
,
<STO_>
and
<On/Off>
. The other keys and
the automatic switch-off function are deactivated.
Energy saving mode
([Eco mode])
If the automatic storing function is active, the meter provides an energy
saving mode ([Eco mode]) to avoid unnecessary energy consumption.
The energy saving mode switches off functions of the meter that are not
required for the automatic storage of measurement data (such as the
display). By pressing any key the energy saving mode is switched off
again.
Duration
1 min ... x min
Storage duration.
Specifies after which time
the automatic storage
should be terminated.
The lower limit of the
storage duration is
restricted by the storage
interval.
The upper limit is restricted
by the number of free
storage locations.
Menu item
Possible
setting
Description
pH
6.949
25.0
°C
0d03h00min
22.09.2008
08:00
0
14
Remaining storage duration
Graphical display of the
storage duration
Eco mode
USB output