![WORLD OF MEDICINE A127 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/world-of-medicine/a127/a127_instructions-for-use-manual_997764054.webp)
Error and Warning Messages
54 / 62
14
Error and Warning Messages
Informational signals
Optical
Acoustic
Priority
Cause
000 (for 3 s)
1 low volume beep
Information
Disposable/reusable tube was
accepted
number of remaining cycles: 0
xxx (for 3 s)
1 low volume beep
Information
Disposable/reusable tube was
accepted
Number of remaining cycles:
xxx
n/a
1 low volume beep
Information
Self-test successfully completed
n/a
1 low volume beep
Information
Key confirmation
Physiological warnings
Optical
Acoustic
Priority
Cause
Overpressure indicator
(flashing, yellow), immedi-
ately until remedied
3 beeps
(after 2 s, alarm condition pre-
vails until remedied)
Medium
Safety threshold for pressure
(200 mmHg) exceeded
Overpressure indicator
(flashing, red), immedi-
ately
until
condition
remedied
10 beeps
(after 0.5 s, alarm condition pre-
vails until remedied)
High
Critical safety threshold for
pressure (250 mmHg) exceeded
OR
Reversing the roller wheel did
not reduce the overpressure
Technical warnings – user error
Optical
Acoustic
Priority
Cause
---
(constant until tube set is
removed)
1 beep
Low
Device was switched on with
tube set inserted, the sensor
could not be checked
E01
(3 s)
1 beep
Low
Attempted pump activation
without tube set inserted
E02
(constant until tube set is
removed)
1 beep
Low
Invalidated tube set has been
inserted
E03
(constant until tube set is
removed)
1 beep
Low
Invalidated tube set has been
inserted
E04 (3 s)
1 beep
Low
Tube set incorrectly inserted
E02/E03
(3x flashing)
1 beep
Medium
Attempted pump activation
with invalid (invalidated or not
approved) tube set inserted
Technical warnings – device error
Optical
Acoustic
Priority
Cause
E05
service icon, red
(constant)
5 beeps
High
Sensor error
E06
service icon, red
(constant)
5 beeps
High
Motor error
E07
service icon, red
(constant)
5 beeps
High
RFID module defective
en
Содержание A127
Страница 1: ...A127 Pump for Arthroscopy A127 Instructions for Use en...
Страница 56: ...Appendix 56 62 15 Appendix 15 1 Test log Date Result Comment Signature en...