Device Setup
20 / 62
CAUTION!
Mains connection
• Make sure the available mains voltage matches the data listed on the type la-
bel attached to the back of the device. Incorrect voltage can cause errors and
malfunctions and may destroy the device.
• Make sure the connection data and technical specifications of the power sup-
ply comply with DIN VDE or national requirements. The mains connection
cable may be plugged only into a properly installed, grounded safety wall
socket (shockproof socket) (see DIN VDE 0100-710).
• Read the device label located in rear of device (type plate) to determine the op-
erating voltage of the device.
Grounding contact
The power connection must be equipped with a grounding contact. Use the ori-
ginal power cable to establish a connection between the mains wall socket and the
non-heating device plug located in the rear of the device.
The grounded, shockproof safety wall socket should be near the device and within
easy reach. Disconnect the device from the mains power supply (pull cable out off
the grounded safety wall socket) if the device is not being used for several days or
an extended period. The device is ready for use as soon as all connections are es-
tablished and all cables have been plugged in.
Only for U.S. operators
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 and
IEC 60320-C13. Grounding will only be reliable if the equipment is connected to a
corresponding hospital grade socket.
Potential equalization
The equipotential bonding is used as a protective measure against the failure of
the protective conductor according to requirements of IEC 60601-1 in the respect-
ively valid version. The installation must be according to the relevant local safety
regulations.
4.3
Mounting the Pump
The device is designed to be suspended from a portable stand or upright placement
in a device rack. Installation and initial startup of the device are to be carried out by
a correspondingly trained service technician authorized by the manufacturer. The
device is installed horizontally with the upper side of the device pointing up. The
height of the installation position of the device above the floor should correspond
with height of the joint.
4.3.1
Mounting on Stand
CAUTION!
Read instructions
Please read the instruction manual for the stand before mounting the pump on the
stand.
CAUTION!
Stability
The position of the device on the stand can affect the stand's stability. If the device
is mounted too high on the stand, the stand may tilt or topple. The user/operator
of the device is responsible for making sure the device cannot be tipped over or
otherwise become unstable.