Using the Device during Surgery
35 / 62
8
Using the Device during Surgery
WARNING!
Functional test
The functional test must be performed by the user prior to each surgery.
WARNING!
Irrigation fluid
The physician must determinate a distension fluid suitable for the application and
medical procedure.
Before surgery
Make sure the device is at the same height as the instrument. A difference in
height affects the pressure readings due to hydrostatic pressure.
Fig. 8–1 Before surgery
1.
Select the desired nominal pressure.
2.
Open the inflow valve at the instrument and close the outflow valve (drain
valve).
3.
Press the START/STOP key. The START operating mode LED (Fig. Front of the
device [
(3)
) lights up. Wait until the instrument recognition is finished
(the instrument is outside of the joint and at a working height of ± 10 cm of
the operation height.
4.
Close the inflow valve after the instrument recognition is finished.
5.
Insert instrument into joint.
6.
Open the inflow valve on the instrument.
7.
The device ensures the joint is distended according to the preset nominal pres-
sure. The flow is automatically reduced once the nominal pressure has been
reached.
After surgery
WARNING!
Emptying the tube set
The manufacturer recommends emptying the tube set using the pump when em-
ploying saccharine irrigation fluids (HF use). Remove fluid bag and instrument first.
1.
Stop the irrigation process.
2.
Close the valve on the instrument. Press the START/STOP key.
3.
User the power switch to turn device off.
4.
Remove standard tube set.