Device Setup
18 / 62
WARNING!
Falls and crashes
Place the device on a stable and level surface. Cables must be laid safely. Tubes
between the device and the patient must not create any obstruction.
WARNING!
Condensation / Water penetration
Protect device from moisture. Do not use if moisture has penetrated the device.
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not operate the device in the vicinity of ex-
plosive anesthetic gases and not in the vicinity of oxygen-enriched environments.
WARNING!
Functional test
The functional test must be performed by the user prior to each surgery.
WARNING!
Original accessories
For your own safety and that of your patient, use only original accessories.
WARNING!
ME System (Medical Electrical System)
Use only parts and/or devices from ME systems (see Chapter Glossary) in patient
environments in compliance with the standard IEC60601-1 in the respective cur-
rently valid version.
CAUTION!
ME systems
The medical electrical (ME) device is suitable for integration in ME equipment sys-
tems (see Chapter Glossary). Operation of the ME device in vicinity of non-ME
devices may result in voiding the intended use of the ME device.
CAUTION!
ME Device in Rack
The ME device may not be used when stacked or stored directly adjacent to or with
other devices. If it should be necessary to operate the device close to stacked with
other devices, the ME device or ME system should be monitored to ensure it works
properly as configured.
CAUTION!
Switching on the device
Before switching on the device, sufficient time must have passed to adapt to the
room climate.