Operating the Device
24 / 62
WARNING!
Reprocessing of sterile disposable products
Infection hazard for patients and/or users and impairment of product functionality
due to reuse. Risk of injury, illness or death due to contamination and/or impaired
functionality of the product! Do not reprocess the product.
CAUTION!
Check of the reprocessed tube set
After each cleaning / reprocessing, the cleaned / reprocessed tube set must be
checked for possible damages caused by the process before it is stored / used prop-
erly and professionally for the following use!
CAUTION!
Pressure chamber diaphragm
The pressure chamber diaphragm is fragile and can be replaced if damaged. The re-
usable tube sets (NOT FOR SALE IN USA) must be prepared before each use. See
Chapter Care of the Reusable Tube Set (NOT FOR SALE IN USA) [
39].
NOTE!
Tube set storage
The tube sets must be stored at room temperature. The shelf life for tube sets is 5
years.
NOTE!
Hygiene regulations
Observe applicable hygiene regulations when disposing of the tube set.
5.6
Using the Irrigation Tube Sets
The A127 is designed to be used with a disposable tube set as well as a reusable
tube set (NOT FOR SALE IN USA).
RFID technology (transponder techno-
logy)
All irrigation tube sets are equipped with a transponder. The transponder techno-
logy detects the tube type, the validity and reliability of a tube set automatically
and a corresponding message is output in the display of the device. This eliminates
virtually all "operating errors" since non fitting, invalid, and not approved tube sets
are reliably detected and not released for use. The corresponding transponder is
located in every approved tube set (see Chapter Irrigation Tube Sets Overview
[
23]).
If a valid tube set has been inserted into the device ready to be operated, a low
volume acoustic signal is emitted and the operating states and warnings display
(Fig. Front of the device [
(5)
) depicts for 3 seconds the number of cycles re-
maining after the current surgery (000 for disposable tube set and for reusable
tube set during last remaining cycle, xxx for reusable tube set).
Invalidating a tube set
If an approved and valid tube set is located in the device, the transponder techno-
logy automatically invalidates this tube set when the device is started. Depending
on the tube set, this has the following consequences:
• Disposable tube set:
Display 000 (3 s) -> remaining cycles after this use
– The tube set can be removed/inserted as often as desired without invalid-
ating the tube set as long as the pump is not started.
– The tube set is invalidated 1 minute after starting the pump.