11
6. Operation
6.1 Emergency Stop systems
Caution – Possible danger of injury or damage.
In case of danger (e.g. when a patient falls down), the treadmill
must be stopped immediately with the Emergency Stop system.
Caution – Possible danger of injury or damage.
Each time the Emergency Stop switch is activated, the main power
is isolated from the system, thereby causing an Emergency Stop.
6.1.1 Emergency Stop pull wire magnetic switch
The Emergency Stop pull wire magnetic switch system consists of a proximity switch that is mounted either to the
WUS data display or else to the handrail and is actuated by means of a magnet. The magnet is connected to a
wire that is attached to the runner's clothing by means of a clip.
When the magnet is pulled away from the proximity switch contact area, the 230 V power supply is interrupted
and an Emergency Stop is initiated.
The magnet must be positioned on the red field with the yellow ring!
When the magnet is not correctly positioned, the treadmill cannot operate!
Caution – Possible danger of injury or damage.
The Emergency Stop pull wire switch magnet must be attached to the runner by means of the
wire with the clip at all times during operation.
6.1.2 Handrail-mounted Emergency Stop switch (mushroom)
The Emergency Stop switch (mushroom type), which is mounted on the side of the right handrail, can be pressed
in case of emergency. When the Emergency Stop switch is pressed, the 230 V power supply is interrupted and an
Emergency Stop is initiated.
To release the Emergency Stop switch, lift the red mushroom button up (pull hard)!
Dismounting in case of emergency
Caution – Possible danger of injury or damage.
To dismount in case of emergency, straddle the treadmill by jumping onto the side panels and
then shut the treadmill off.