28
The device is working
properly, but the pump is
being forced.
Too much pressure by the
impeller on the tube in the
peristaltic pump.
Check that the tube in
the peristaltic pump has
been correctly inserted.
The pump is running
correctly but when it
stops liquid comes out of
the handpiece.
The door of the peristaltic
p u m p i s n o t c l o s e d
properly.
Make sure that the door
of the peristaltic pump is
properly closed.
Insufficient power.
The tip is not correctly
fitted to the handpiece (the
message WARN appears on
the display).
Unscrew the tip and
screw it back into place
correctly.
The tip is worn, broken
or deformed (the message
WARN appears on the
display).
Replace the tip.
LCD screen mess or
imcomplete display.
Voltage interference.
Stop any operation,
change the model then
return to the original
model or restart the
machine.
15. Technical data
15.1 Device in accordance with Directive 93/42EEC.
15.2 According to EN60529: IPX1 (device)
Содержание ULTRASURGERY