Problem
Possible cause
solution
The device does not turn
on when the switch it
positioned on ON.
The connector on the end of
the power cable is plugged
into the socket on the rear
of the device properly.
Check that the power
cable is firmly connected.
The power cable is faulty. Check that the power
outlet is working
properly. Replace the
power cable.
The fuses blew out.
Replace the fuses.
The connector on the
end of the power cable is
plugged into the socket
on the rear of the device
properly.
The connector of the
footswitch is not properly
plugged into the socket.
Insert the footswitch
connector
properly.
The footswitch will not
work.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
A faint whistle can be
heard coming from the
ULTRASURGERY
handpiece during
operation.
The tip is not correctly
tightened onto the
handpiece.
Unscrew the tip and
screw it back into place
correctly.
The device is switched
on but does
not work, the message
WARN appears on the
display.
The tip is not fitted correctly
into the handpiece.
Unscrew the tip and
screw it back into place
correctly.
The tip is worn, broken or
deformed.
Replace the tip.
The connector of the cord
is wet.
Dry the connectors.