27
Problem
Possible cause
solution
The device is switched
on but will not work, the
message WARN appears
on the display.
Cord not connected to the
device.
Connect the cord to the
device.
Lack of continuity of a lead
in the cord.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
Handpiece failure.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
Malfunctioning of the
tuning circuit.
Contact the nearest
dealer or authorized
Woodpecker service
centre.
No liquid comes out of
the tip during operation.
The tip is of the type with
no through-flow of liquid.
Use an tip of the type
with through-flow of
liquid.
The bag of liquid is empty. Replace the bag with a
full one.
The cover of pump that
connected with the water
tube is open.
Close the cover.
The tubes of the drip system
and of the pump have not
been correctly installed.
Check the connections of
the tubes.
The tip is clogged.
Free the passage in the
tip through which the
water passes.