![Woodpecker ULTRASURGERY Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/woodpecker/ultrasurgery/ultrasurgery_manual-of-use-and-maintenance_995151025.webp)
22
9.1 handle this device gently, keep away from the shake source, and should
install and store in shade.
9.2 Do not mix with poison, causticity, explosive and combustible things
together.
9.3 store in the environment of humidity under 80%, pressure around: 50kPa
~
106kPa, temperature between -10
℃ ~
+50
℃
.
9.4 If the device is not used for a long time running, it is better to connect the
electricity and water one time per month, 5 minutes per time.
9.5 disconnect the device from the power mains.
Danger: Check regularly that the power cable is intact, if it is
damaged, replace it with an Woodpecker spare.
10. Replacement of the fuses
Danger: Switch off the apparatus.
Always turn off the apparatus by means of the switch (Fig.5-Ref.B) and
disconnect it from the power outlet before carrying out the following maintenance
activities.
10.1 Insert the flat tip of a screwdriver into the recess in the fuse compartment
below the power socket and use it as a lever (Fig.12-Ref.A);
10.2 Pull out the fuse compartment(Fig.12-Ref.B);
10.3 Danger: Replace the fuses, using fuses of the type indicated on the
identification label on the bottom of the apparatus;