16
Warning code
Possible Cause
Warn 05
Beyond the electricity.
Warn 06
Tip is not correctly secured to the handpiece
or tip is worn or broken or deformed.
Handpiece failure.
Warn 07
Tuning circuit not working properly.
Please restart the device, if the problem
persists, stop using it and call the
Woodpecker service centre immediately.
7.5 Instruction for use
7.5.1 Open the air intake on the drip system;
7.5.2 Screw the chosen tip onto the ULTRASURGERY handpiece until it is flush
against it;
7.5.3 To use the torque wrench correctly (Fig.8) proceed as follow;
a) Hold the body of the handpiece firmly;
Warning: Do not grip the end part of the handpiece or the cord, only
the plastic casting (Fig.8) and do not turn it while fastening the tip in place;
b) Turn the wrench in a clockwise direction until the cultch engages (till making
clicking sound);
c) The tip is now properly tightened in place;
7.5.4 Make sure that the ULTRASURGERY handpiece is correctly connected to
the handpiece connector (Fig.6-Ref.C, Fig.7);
7.5.5 Check the display to see the type of power that has been set. If the type of
power required different from the type that has been set, use key “M” (see 7.1) to
switch;
7.5.6 Check the display to see the power level that has been set, if the type of
power required differs from the level that has been set, use the key “+”/“
-
”(Fig.10)
for selecting, depending on the type of function that has been set;
7.5.7 Check the display to see the delivery rate of the peristaltic pump, if the
delivery rate required is other than the level that has been set, use the key