d) Special: Low bone density
7.2.2 Root function (Fig.10)
In this function, both the water and power model are available, two models as
follow:
a) Perio.
b) Endo.
7.2.3 Clean function (Fig.11)
In this function, press the footswitch and hold it on for 3 seconds, the device can
clean the tube automatically in 25 seconds.
7.3 Safety requirements during use.
Danger: Contraindications.
Do not use the ULTRASURGERY on patients with pacemakers or other
implantable electronic devices. This requirement also applies to the operator.
Danger: Breakage and wear of the tips.
The high-frequency vibrations and wear may, very occasionally, lead to breakage
of the tip. Tips of which the shape has been changed or which are otherwise
damaged are liable to break during use. Any such tips should definitely not be
used. It is necessary to instruct the patient to breathe through his nose during the
treatment in order to avoid ingestion of the broken fragment of the tip.
Danger: Control of infections.
For maximum safety of both the patient and the operator, clean, disinfect and
sterilize the piezo electronic handpiece, the tips and the torque wrench after each
treatment.
Warning: Contraindication.
Do not carry out this treatment on metal or ceramic prosthetic artifacts. The
ultrasonic vibrations could lead to decrementing of such artifacts.
Warning: Contraindication.