After autoclave sterilizing of the handpiece, wait for it to cool down completely
before using it.
Warning: The electrical contacts inside the cord connector must be
dry.
Before connecting the handpiece to the device, make sure the electrical contacts
of the connector are perfectly dry, in particular after the autoclave sterilization
cycle. If necessary, dry the contacts by blowing air onto them with the syringe.
Warning: To use the device correctly, it is necessary to press
the footswitch and start it up without letting the tip rest on the part to
be treated. This will allow the electronic circuit to detect the point where
resonance of the tip is without any interference, thus enabling optimum
performance.
If this is not done, contact with the part to be treated or with other surfaces before
start-up could cause tripping of the protection systems.
Warning: For spray treatment, use only tips through which liquid
is passed.
7.4 Protection systems and alarms.
The device has a diagnostics circuit that is used to recognize tripping of the
protection system and of the alarms. These are shown on the display, as follows:
Warning code
Possible Cause
Warn 01
There is a signal transmission problem.
Switch the device off and then on again, If
the problem persists contact Woodpecker
service centre.
Warn 02
Tuning circuit not working properly.
Handpiece failure.
Warn 03
Fan failure.
Warn 04
Pump failure.