8. Cleaning, Disinfection and Sterilization.
8.1CLEAN function—Cleaning of the liquid circuit
Warning: Failure to carry out cleaning of the tubes will lead to
crystallization of salts that can seriously damage the equipment.
Warning: Handpiece and cord can’t be detached.
a) Change the bag containing water (demineralized water is recommended);
b) Check whether the water system is connected correctly;
c) Start the CLEAN function (Fig.11);
d) Press the footswitch to start the cleaning cycle, as soon as the peristaltic pump
starts up, a status bar will appear on the display to indicate progressively the
time remaining to completion of the tempo CLEAN cycle. The cycle lasts for 25
seconds and cannot be stopped;
e) Once the cleaning cycle has been completed, the device exits from the CLEAN
function and returns to BONE function (Fig.9);
f) On completion of the cleaning operations, empty the tubes and dry the
accessories that have been through the cleaning cycle.
8.2 Cleaning and disinfecting the casing of the apparatus
Danger: The casing of the apparatus is not protected against the
penetration of liquids. Do not spray liquids directly onto the surface of the
casing of the apparatus.
Danger: The apparatus cannot be sterilized.
After each treatment carry out the following operations:
a) Remove the tip from the handpiece;
b) Clean and disinfect the surfaces of the casting, the cords and their connectors
using a cloth moistened with a mild detergent or disinfectant solution with a
neutral PH (ph 7). Follow carefully the instructions given by the manufacturer of