Guidance and manufacturer's declaration - electromagnetic immunity (2/2)
The 6001 is intended for use in the electromagnetic environment specified below. The customer or the
user of the 6001 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Portable
and
mobile
RF
communications equipment should
be used no closer to any part of the
1608A, including cables, than the
recommended separation distance
calculated
from
the
equation
applicable to the frequency of the
transmitter.
Recommended
separation
distance:
Conducted
RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = [3,5/V1] x
P = 1,2 x
P
Radiated RF IEC
61000-4-3
3 V/m
80 MHz to 2,5 GHz
3 V/m
At 80 bis 800 MHz:
d = [3,5/E1] x
P = 1,2 x
P
At 800 MHz bis 2,5 GHz:
d = [7/V1] x
P = 2,3 x
P
where P is the maximum output
power rating of the transmitter in
watts
(W)
according
to
the
transmitter manufacturer and d is
the
recommended
separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as deter-mined by an
electromagnetic site survey, a
should be less than the compli-
ance level in each frequency
range. b
Interference may occur in the
vicinity of equipment marked with