GA_Portio-Koagulator_E_150914.docx
Page 32 of 44
7
T
ROUBLE SHOOTING
Fault Description
Possible Causes
Remedy
No device function
No display is activated
Mains switch
1
does not
light up
Failure at mains supply
Defective mains fuses
Have supply network
checked
Exchange fuses
(observe type of fuse,
see „Technical Data)
Have device serviced
by manufacturer or by
authorized dealer
Mains switch
1
illuminates, no device
function
Internal device failure
Have device serviced
by manufacturer or
authorized dealer
Device does not heat up
control lamp, RED:
3
illuminates
Heating is electronically
interrupted
therapy probe
3
is
defective or not plugged
in.
Temperature is too
high.
Internal device failure.
Turn device off, connect
new probe and turn
device on again
Have device serviced
by manufacturer or
authorized dealer
CAUTION!
If the probe comes into contact with tissue, no sizzling sound may be
audible. Such sound would signalize exceeded coagulation temperature
(above 100°C).