4.4 ASSEMBLING
4.4.1 Connection of Instrument Cable 3 and Therapy Probe 4
Take the therapy probe
(4)
and join with receptacle
(2)
of the unit through instrument
cable
(3)
.
Rest the Therapy Probe in one of the fixtures for the therapy probes
(11)
and make sure
the tip is not in contact with cloth or any other material.
CAUTION!
In case of obvious or suspected defects in the instrument cable and / or the
therapy probe, this must be replaced immediately.
CAUTION!
Defective instrument cable and therapy probes can be sent for repair to the
WISAP Medical Technology GmbH or to an approved contractor.
4.5 FIRST PUTTING INTO OPERATION
CAUTION!
Please make sure that no liquid enters into the Controller unit or that the
unit does not get in touch with liquids.
CAUTION!
In order to ensure good dissipation of heat generated during an operation
the control unit must not be covered with cloth.
NOTE!
Install the device onto a plane surface.
NOTE!
For the correct use of this device it is important that the device adapts to
the ambient conditions (room temperature). Please wait about 60 minutes
after installation before you begin the application. This applies both to the
first use and in the event that the device is transported to another
environment.