GA_Portio-Koagulator_E_150914.docx
Page 23 of 44
4.5.1 Connection to the equipotential bonding
Ground the units by connecting a bonding conductor with the plug-in connector
(14)
(according to DIN 42801).
The potential equalization connection
15
of Cervix Coagulator 6001 has to be connected
via a suitable grounding cable to the central potential equalization or if installed in / on
a trolley with the potential equalization connection of the trolley.
DANGER!
The device must never be operated with a defective power cord.
WARNING!
The device is not destined for operation in explosive-endangered areas.
WARNING!
If explosive ANESTHETICS are in
use, device and accessories must
not be operated in the highlighted
zone.
CAUTION!
Run a visual inspection of this device before each use. When detecting
damage set the unit aside immediately and do not perform any operation
with it. Contact the manufacturer immediately.
CAUTION!
Check the appliance and all accessories for proper operation before each
procedure. In case of detected or suspected defects the products is not to
be used
4.5.2 Connection to the power supply
Check before first use, whether the local mains voltage matches the voltage specified
on the product label
(12).
Only then can you connect your device to the mains power
supply.
Plug the power cord into the connector
(13)
on the back of the Cervix Coagulator 6001.
Before mains connection of Cervix Coagulator 6001 bring the mains power switch
(1)
in
position "0" = OFF. Connect the power cord to the power connector
(13)
, and then
connect to the mains.