Plug the power cord into the connector
(13)
at the rear of Cervix Coagulator 6001.
CAUTION!
The device unit may only be installed by the manufacturer or by
AUTHORIZED PERSONNEL.
CAUTION!
The device may only be operated in medically utilized rooms, which have
been installed according to the guidelines of DIN VDE 0107.
CAUTION!
To avoid the risk of electrical shock, this unit must be connected only to a
supply with protective conductor:
CAUTION!
If the Cervic Coagulator is connected into a system with other electrical
equipment or medical devices, the system has to comply with the
requirements of IEC 60601-1.
CAUTION!
Place the unit out of the reach of the patient!
NOTE!
For the correct use of this device it is important that the device adapts to
the ambient conditions (room temperature). Please wait about 60 minutes
after installation before you begin the application. This applies both to the
first use and in the event that the device is transported to another
environment.