Guidance and manufacturer's declaration - electromagnetic immunity (1/2)
The 6001 is intended for use in the electromagnetic environment specified below. The customer or the
user of the 6001 should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical
fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1
kV
for
input/output lines
±2 kV for power
supply lines
±1
kV
for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage
dips,
short
interruptions and voltage
variations
on
power
supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip
in UT) for 0,5 cycle
40 % UT (60 % dip in
UT) for 5 cycles
70 % UT (30 % dip in
UT) for 25 cycles
<5 % UT (>95 % dip
in UT) for 5 sec
Complies
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial
or
hospital
environment.
NOTE: UT is the a. c. mains voltage prior to application of the test level.