Reference Manual
00809-0100-4160, Rev AB
March 2010
2-21
Rosemount 2160
CALIBRATION
Dry Fork Frequency
The 2160 detects the process is Wet or Dry by monitoring the vibrating fork
sensor frequency.
Dry Fork Frequency
is set when calibrated under Dry
process conditions at the factory.
See also “Dry Fork Frequency / Switch Points” on page 2-18 for associated
parameters.
To view the dry fork frequency:
1. From the
Home
screen, select
3: Service Tools
.
2. Select
4: Maintenance
.
3. Select
3: Calibrate
.
4. Select
1: Sensor Calibration
.
5. Select
1: Dry Fork Frequency
.
NOTE
The Fast Key sequence 3, 4, 1, 3, 1 also indicates Dry Fork Frequency.
Sensor Frequency
See “Sensor Frequency” on page 2-16 for a description of this parameter.
Calibrate Dry Fork
This command starts the
on-site re-calibration
of the vibrating fork sensor in
Dry process conditions
. It should only be performed by authorized persons.
A comparison is made between the
new dry fork frequency
and original
factory-set
Dry Fork Frequency
value (page 2-21). If the difference is greater
than
Allowable Change In Dry Fork Frequency
(page 2-12), the
re-calibration is rejected. Check the fork for damage, corrosion, or coating,
and clean the fork if necessary before re-trying.
When the re-calibration is successful,
Dry Fork Frequency
is set to the
new
dry frequency
.
To start the on-site re-calibration:
1. From the
Home
screen, select
3: Service Tools
.
2. Select
4: Maintenance
.
3. Select
3: Calibrate
.
4. Select
1: Sensor Calibration
.
5. Select
3: Calibrate Dry Fork Frequency
.
6. Input the process temperature when prompted.
Sensor Calibration
Status
The 2160 is supplied with a factory calibration but can be re-calibrated on-site
using the
Calibrate Dry Fork
command (page 2-21).
Sensor Calibration Status
indicates the 2160 is:
•
Factory Calibrated
–
No further calibration is normally required
.
•
Site Calibrated
–
Calibration successfully performed on-site.
•
Un-calibrated
–
Calibration is required. Contact the factory
.
Fast Keys
3, 4, 3, 1, 1
Fast Keys
3, 4, 3, 1, 2
Fast Keys
3, 4, 3, 1, 3
Fast Keys
3, 4, 3, 1, 4
Содержание Rosemount 2160 WirelessHART
Страница 2: ......
Страница 18: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 1 10...
Страница 56: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 3 14...
Страница 70: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 5 10...
Страница 76: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 6 6...
Страница 92: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 A 16...
Страница 98: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 B 6...
Страница 99: ......