Reference Manual
00809-0100-4160, Rev AB
March 2010
Rosemount 2160
2-8
CHECK OUTPUT
Before performing other level switch operations, ensure the 2160 is operating
properly by checking the process variables.
Variables
To view the Variables menu:
1. From the
Home
screen, select
3: Service Tools
.
2. Select
2: Variables
.
The variable menu displays the following process variables:
•
PV
–
level switch output state of
0.0
(Dry) or
1.0
(Wet)
•
PV Status
–
valid or fault status (see “Troubleshooting” on page 6-2)
•
Sensor Frequency
–
the frequency of the vibrating fork
•
Sensor Frequency State
–
valid or fault status
•
Electronics Temperature
–
the temperature inside the 2160 housing
•
Electronics Temperature Status
–
valid or fault status
•
Supply Voltage
–
the input voltage to the level switch
•
Supply Voltage Status
–
valid or fault status
•
Last Update Time
–
elapsed
time since the last data update
BASIC SETUP
Temperature Units
The
Unit
setting allows the
Electronics Temperature
process variable to be
displayed in Farenheit (°F) or Celsius (°C).
To view the electronics temperature menu:
1. From the
Home
screen, select
2: Configure
.
2. Select
2: Manual Setup
.
3. Select
3: Device Temperature
.
4. Select
1: Electronics Temperature
.
The electronics temperature menu displays the following:
•
Unit
–
Change/View measurement units for
Electronics Temperature
•
Maximum
–
View the highest measured electronics temperature
•
Minimum
–
View the lowest measured electronics temperature
Operation Mode
The 2160 has three operation modes:
•
Standard –
Default operation mode for no sensor fault detection
•
Enhanced (Fault=WET) –
Output is forced to Wet when fault detected
(see Figure A-2 on page A-3)
•
Enhanced (Fault=DRY) –
Output is forced to Dry when fault detected
(see Figure A-2 on page A-3)
To change or view the operation mode:
1. From the
Home
screen, select
2: Configure
.
2. Select
2: Manual Setup
.
3. Select
2: Operation
.
4. Select
1: Application
.
5. Select
1: Operation Mode
.
Fast Keys
3, 2
Fast Keys
2, 2, 3, 1
Fast Keys
2, 2, 2, 1, 1
Содержание Rosemount 2160 WirelessHART
Страница 2: ......
Страница 18: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 1 10...
Страница 56: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 3 14...
Страница 70: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 5 10...
Страница 76: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 6 6...
Страница 92: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 A 16...
Страница 98: ...Reference Manual 00809 0100 4160 Rev AB March 2010 Rosemount 2160 B 6...
Страница 99: ......