Resynchronizing after Parameter Changes
With the exception of the number-of-steps parame-
ter, only parameters changed on the Slave 681XXA
during master-slave operation in step sweep fre-
quency mode cause the swept frequency output sig-
nals to become unsynchronized.
To resynchronize the swept frequency output sig-
nals after making a parameter change on the Slave
681XXA, first put the Master 681XXA in CW mode,
then place it back in step sweep mode.
If you change the number of steps in the step sweep
on either sweep generator, you must select the same
number of steps on the other sweep generator. To re-
start/resynchronize master-slave operation, put the
Master 681XXA in CW mode, then place it back in
step sweep mode.
LOCAL (FRONT
MASTER-SLAVE
PANEL) OPERATION
OPERATION
3-72
681XXA OM