POWER LEVEL SWITCHING TIME (to within specified
accuracy):
Without Change in Step Attenuator:
<50
µ
s
With Change in Step Attenuator:
<20 ms
ACCURACY AND FLATNESS
Step Sweep and CW Modes
Accuracy:
±
1.0 dB
Flatness:
±
0.8 dB
Analog Sweep Mode (typical)
OTHER OUTPUT POWER SPECIFICATIONS
Source Impedance: 50
Ω
nomimal
Source SWR (Internal Leveling):
Without Attenuator:
<1.7 at <2 GHz typical
<1.6 at 2 to 20 GHz typical
<2.0 at >20 GHz typical
With Attenuator:
<2.0 typical
Output Power Resolution: 0.01 dB
Power Level Stability with Temperature:
0.02 dB/°C typical
Level Offset: Offsets the displayed power level to establish
a new reference level.
OUTPUT ON/OFF: Toggles the RF output between an Off
and On state. During the Off state, the RF oscillator is turned
off. The On or Off state is indicated by two LEDs located
below the OUTPUT ON/OFF key on the front panel.
RF On/Off Between Frequency Steps: System menu se-
lection of RF On or RF Off during frequency switching in CW
or Step Sweep modes.
RF On/Off During Retrace: System menu selection of RF
On or RF Off during retrace.
Internal Leveling: Power is leveled at the output connector
in all modes.
External Leveling:
External Detector:
Levels output power at a remote de-
tector location. Accepts a positive or negative 0.5 mV to
500 mV input signal from the remote detector. EXT ALC
GAIN adjusts the input signal range to an optimum value.
BNC connector, front and rear panel.
External Power Meter:
Levels output power at a re-
mote power meter location. Accepts a
±
1V full scale input
signal from the remote power meter. EXT ALC GAIN ad-
justs the input signal range to an optimum value. BNC
connector, front and rear panel.
External Leveling Bandwidth:
30 kHz typical in Detector mode.
0.7 Hz typical in Power Meter mode.
CW POWER SWEEP
Range: Sweeps between any two power levels at a single
CW frequency.
Resolution: 0.01 dB/step
Accuracy: Same as CW power accuracy.
Step Size: User-controlled, 0.01 dB to the full power range
of the instrument.
Step Dwell Time: Variable from 1 ms to 99 seconds. If the
sweep crosses a step attenuator setting, there will be a
sweep dwell of approximately 20 ms to allow setting of the
step attenuator.
SWEEP FREQUENCY/STEP POWER
A power level step occurs after each frequency sweep.
Power level remains constant for the length of time required
to complete each sweep.
Attenuation
Below
Max Power
Frequency
(GHz)
0.01-20
20-40
Accuracy
0-12 dB
±
1.0 dB
±
2.0 dB
12-30 dB
±
3.5 dB
±
4.6 dB
30-60 dB
±
4.0 dB
±
5.2 dB
60-122 dB
±
5.0 dB
±
6.2 dB
Flatness
0-12 dB
±
1.0 dB
±
2.0 dB
12-30 dB
±
3.0 dB
±
4.1 dB
30-60 dB
±
3.5 dB
±
4.6 dB
60-122 dB
±
4.0 dB
±
5.2 dB
PERFORMANCE
SPECIFICATIONS
681XXA OM
B-3