Power Level
Flatness Test
Procedure
Power level flatness is checked by measuring the
power level variation during a full band sweep; first
in the step sweep mode, then in the analog sweep
mode.
1. Set up the 681XXA as follows for a step sweep
power level flatness test:
a. Reset the instrument by pressing
SYSTEM
,
then
Reset
. The CW Menu is displayed.
b. Press
Step
to place the 681XXA in the step
sweep frequency mode and display the Step
Sweep Menu (shown below).
c. With the Step Sweep menu displayed, press the
main menu key
FREQUENCY
CONTROL
The Sweep Frequency Control menu, shown be-
low, is displayed.
d.Press
Full
to select a full range frequency
sweep.
e. Press
Edit L1
to open the current power level
parameter for editing.
f. Set L1 to the power level noted on the test re-
cord.
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
681XXA OM
5-11