g. Now, return to the Step Sweep menu by press-
ing the main menu key
CW/SWEEP
SELECT
h. At the Step Sweep menu, press
Sweep Ramp
to go to the Step Sweep Ramp menu (shown be-
low).
i. Press
Dwell Time
to open the dwell time-per-
step parameter for editing.
j. Set the dwell time to 1 second.
NOTE
Monitor the 681XXA’s Horizontal Output on
the Oscilloscope to determine sweep start
and stop.
2. As the 681XXA steps through the full frequency
range, measure the maximum and minimum
Power Meter readings and record the values on
the test record. Verify that the variation (differ-
ence between the maximum and minimum read-
ings) does not exceed the value noted on the test
record.
OPERATION
POWER LEVEL ACCURACY
VERIFICATION
AND FLATNESS TESTS
5-12
681XXA OM