UNLEVELED
is Displayed
Step 1.
Check that the output power does not exceed the specified lev-
eled-power rating and that the
RF OUTPUT
connector is ter-
minated into a 50
Ω
load.
Reduce the power level to not exceed the specified leveled-
power rating or terminate the RF OUTPUT connector
with a 50
Ω
load.
If error message remains displayed, call a service techni-
cian.
RANGE ERR
is Displayed
Step 1.
Check that (1) the sweep start frequency entered is not
greater than the stop frequency, (2) the dF value entered does
not try to set the frequency sweep outside the range of the
sweep generator, or (3) the step size entered is not greater
than F2 minus F1.
Enter a valid sweep start frequency, dF value, or step size.
If the error message remains displayed, call a service tech-
nician.
Table 6-4.
Troubleshooting (3 of 3)
OPERATOR
MAINTENANCE
TROUBLESHOOTING
6-12
681XXA OM