15. Specifications
Operating conditions
Range of applications
Indoor and outdoor application
Direct exposure to sunlight is permitted.
Humidity
≤ 93 % r. h.
Ignition protection types
See product label
II 1/2G Ex ia IIC T4/T5/T6 Ga/Gb
II 2G Ex ia IIC T4/T5/T6 Gb
II 3G Ex ic IIC T4/T5/T6 Gc
II 1/2D Ex ia IIIC T135°C Da/Db
II 2D Ex ia IIIC T135°C Db
Vibration resistance
4 g (5 ... 100 Hz) per GL characteristic curve 2
Shock resistance
150 g (3.2 ms) per IEC 60068-2-27
Ingress protection
IP 66/67
IP 65 for versions with circular connector,
angular connector or overvoltage protection
Ingress protection only applies with closed case
head and closed cable glands.
Permissible temperature ranges
Temperature class / surface temperatures for all variants
except for
UPT-2*-*I*-**-**-*****-*4****-** (model without cooling element):
Temperature class / surface temperature
Ambient and medium temperature (°C)
T5, T6
-40 ≤ Ta ≤ +60
T4
-40 ≤ Ta ≤ +80
T135 °C
-40 ≤ Ta ≤ +40 for Pi = 750 mW
-40 ≤ Ta ≤ +70 for Pi = 650 mW
-40 ≤ Ta ≤ +80 for Pi = 550 mW
Temperature classes / surface temperatures
only for
variant
UPT-2*-*I*-**-**-*****-*4****-** (model with cooling element):
Temperature class
Max. medium temperature (°C) Ambient temperature (°C)
T4
120
-40 ≤ Ta ≤ +50
T3
150
-40 ≤ Ta ≤ +40