
2187800.09 GB/D/F/E 04/2008
14
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio IS-10, IS-11
2187800.09 GB/D/F/E 04/2008
15
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio IS-10, IS-11
Specifications
Model IS-10 / IS-11
RF-immunity
V/m
10 {30}
BURST
KV
4
Shock resistance
g
1000 according to IEC 60068-2-27 (mechanical shock)
Vibration resistance
g
20 according to IEC 60068-2-6 (vibration under resonance)
Wiring protection
Protected against reverse polarity and short circuiting on the instru-
ment side
Weight
kg
Approx. 0.2
When designing your plant, take into account that the stated values (e.g.burst pressure,
over pressure safety) apply depending on the material, thread and sealing element used.
7. Starting, operation
GB
*)
In an oxygen version model IS-11 is not available. In an oxygen version model IS-10 is only available in gauge pressure ranges
≥
0.25 bar with media temperatures between -4 … +140 °F / -20 … +60 °C and using stainless steel or Elgiloy
®
wetted parts.
{ } Items in curved brackets are optional extras for additional price.
Functional test
Open pressure connections only after the system is without pressure!
Observe the ambient and working conditions outlined in section 7
„Technical data.
Please make sure that the pressure transmitter is only used within the over
load threshold limit at all times!
When touching the pressure transmitter, keep in mind that the surfaces of
the instrument components might get hot during operation.
The output signal must be proportional to the pressure. If not, this might point to a
damage of the diaphragm. In that case refer to chapter 10 „Troubleshooting“.
Z = Zero
S = Span
Female
connector
Sealing
Clamping
nut
Male
connector
Sealing
Case with
pressure
connection
Recommended recalibration cycle: 1 year
For further information (+49) 9372/132-295
8. Adjustment of zero point / span (only for pressure transmitter with
clamping nut)
A
B
We do not recommend to adjust the span potentiometer. It is used for adjustment ex
factory and should not be adjusted by you unless you have adequate calibration equip-
ment at your disposal (at least three times more accurate than the instrument being
tested).
8. Adjustment of zero point / span
GB
Make sure wires are not cut or pinche
during disassembly and reassembly of
the connector.
Remove the female connector. Open
the pressure transmitter by detaching
the clamping nut (see Fig. A ). Carefully
remove the male connector from the
case.
Adjust the zero point (Z) (see Fig. B ) by
generating the lower limit of the pressure
range.
Adjust the span (S) by generating the
higher limit of the pressure range.
Check the zero point.
If the zero point is incorrect, repeat
procedure as required.
Reassemble the instrument carefully.
Make sure all sealings and o-rings are
not damaged and correctly installed to
assure the rated moisture ingress protec-
tion.
Warning
Caution