Error

Possible cause

Remedy

- - - -

Interruption of the internal reference 
sensor or the internal reference 
sensor is defective.

Servicing required:
Please return the instrument to 
the manufacturer or an authorised 
service partner

uuuu

Measured temperature under the limit 
value of the internal reference sensor 
(under range -200 °C)

oooo

Measured temperature above the 
limit value of the internal reference 
sensor(over range +850 °C)

ErEP

Possible fault in the EEPROM 
memory of the controller

Press the P key

Fan not running

The fan is defective or blocked

Servicing required:
The temperature switch is possibly 
triggered, switching off the power 
supply to the heating cartridge 

End temperature is not 
achieved

Solid state relay is defective or the 
heating/cooling element has short 
circuited or aged

Servicing required

No display

Controller defective

Servicing required

No function

Network connection not established 
correctly or fuse defective

Check the network connection and 
fuse

GB

Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100

11263911  04/2007 GB/D

WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths

7. Error Messages

If servicing is required, shut down the calibrator/
micro calibration bath (see chapter 12) and return it 
to the manufacturer.

8. Cooling Down of the Metal Block/Liquid Bath

Risk of burns!
Prior to transport or contact with 
the metal block/liquid bath ensure 
that it has cooLED down sufficiently; 
otherwise there is a risk of severe 
burns at the metal block/liquid bath 
and the test specimen.
In order to cool down the metal 
block/liquid bath quickly, set the set 
temperature to a low temperature, 
e.g. room temperature. 

The instalLED fan gently and automatically switches 
to a higher speed for heating instruments, thus 
providing more cooling air. The LED OUT 2 indicates 
the status of the output for the fan control. If the 
LED OUT 2 lights up, the fan is running at high 
speed. If the LED OUT 2 does not light up, the fan is 
running at low speed. 

The controller switches the active cooling on for 
heating/cooling instruments. The LED OUT 2 
indicates the status of the output for the active 
cooling. If the LED OUT 2 lights up, the active 
cooling is running. If the LED OUT 2 does not light 
up, the cooling is not active. 

After switching off or after removing 
the mains connection, the instalLED 
fan can no longer provide cooling 
air. Nevertheless, sufficient thermal 
isolation between the metal block/
liquid bath and the housing is still 
guaranteed. 

Содержание CTB9100-165

Страница 1: ...Temperature Dry Well Calibrators Micro Calibration Baths Kalibrator temperatury z suchym otworem pomiarowym Kalibrator k pielowy mikro CTD9100 165 450 650 CTB9100 165 225 Kalibrator temperatury z such...

Страница 2: ...on notes and tips Caution General point of hazard Please heed the operating instructions Danger Hazard of personal injury through electric voltage Danger Hazard of personal injury through high tempera...

Страница 3: ...alibration Bath 14 7 Error Messages 25 8 Cooling Down of the Metal Block Liquid Bath 25 9 Cleaning and Maintenance 26 10 Warranty and Repairs 26 11 Recalibrating 26 12 Decommissioning and Disposal 27...

Страница 4: ...ns apply to the following calibrator micro calibration bath types 1 Device Description and Intended Use Temperature calibrators CTD9100 165 cooling and heating CTD9100 450 heating CTD9100 650 heating...

Страница 5: ...ou will find the type label with important information about the individual model CTD9100 165 range 30 C 165 C CTD9100 450 range 40 C 450 C CTD9100 650 range 40 C 650 C The suitable line voltage and f...

Страница 6: ...lt in personal injuries Correct and safe operation of the calibrator micro calibration bath demands correct transport storage installation and assembly as well as proper use and careful operation and...

Страница 7: ...with your eyes Since silicone oil is hygroscopic always use the transport lid to close the calibration bath after use The transport lid is equipped with a safety valve If the micro calibration bath i...

Страница 8: ...calibration bath is delivered in special protective packaging Keep this protective packaging for sending the instrument for recalibration or repairs to the manufacturer Inspect the delivery first Stan...

Страница 9: ...the calibra tor micro calibration bath is not heating 6 LED OUT 2 a Heating instrument Signals the status of the output for the fan control If the LED OUT 2 lights up the fan is running at high speed...

Страница 10: ...externally converted into RS 232 or USB signals Appropriate converters including drivers are optionally available The PC records all the operating data and enables the programming of all the calibrato...

Страница 11: ...emember to clean the sleeve and the block This prevents the sleeves becoming jammed in the heating block Fig 8 Sleeves 5 3 Preparing the micro calibration bath In order to achieve the best possible ac...

Страница 12: ...ating the magnetic stirrer The best possible homogeneity is achieved by stirring the calibration liquid with the magnetic stirrer Set the stirring speed to the respective max speed Turn the thumb whee...

Страница 13: ...ation bath Connect the supplied mains plug to a mains outlet Actuate the mains switch The controller is initialized tESt appears on the upper PV display The version number e g rL 2 2 appears on the lo...

Страница 14: ...bration mode HEATING displays The LED OUT 2 indicates the status of the output for the fan cooling control Fig 16 Calibration mode FAN or COOLING displays a Heating instrument The LED OUT 2 indicates...

Страница 15: ...s Use the 6 and 5 keys to select the desired main menu see overview Press the P key to confirm the selected menu item Fig 18 Menu structure main menu Press the and 6 key to raise and lower the value b...

Страница 16: ...he menu structure it is possible to reach the group and parameter levels to carry out settings via OPEr Returning to another level If no key is pressed in the main menu at the group or parameter level...

Страница 17: ...emperature appears on the lower SV display The control has now been switched off and the reference temperature will constantly drop without being regulated Fig 20 Main menu display Fig 21 Menu control...

Страница 18: ...Main menu display Fig 26 Menu manual control OPLO Fig 27 Manual control OPLO setting display Press the 5 key to increase the output capacity Press the 6 key to decrease the output capacity Press the...

Страница 19: ...e memory The selected set value memory e g SP3 flashes on the upper PV display The corresponding current set temperature appears on the lower SV display Press the 5 key to increase the set temperature...

Страница 20: ...the calibrator micro calibration bath it is automatically determined which of the gradients heating gradient SLor or cooling gradient SLoF is to be used Additionally you can ensure that the calibrator...

Страница 21: ...temperature or switching the calibrator off and on again A further procedure is to switch off the automatic control prior to modifying parameters see section 6 3 1 and to switch it on again afterward...

Страница 22: ...values only when modifying the set temperature or switching the calibrator micro calibration bath off and on again 6 3 7 2 Setting the cooling gradient The cooling gradient SLoF is active if the refe...

Страница 23: ...returns to a previous level up to the calibration mode After carrying out the settings the calibrator micro calibration bath uses the new values only when modifying the set temperature or switching th...

Страница 24: ...second the value increases or decreases quickly and after two seconds even more quickly this means the desired value can be reached rapidly Press the P key to confirm the set duration time dur t The...

Страница 25: ...Baths 7 Error Messages If servicing is required shut down the calibrator micro calibration bath see chapter 12 and return it to the manufacturer 8 Cooling Down of the Metal Block Liquid Bath Risk of b...

Страница 26: ...the outside of the calibrator micro calibration bath with a damp cloth and some water or with a solvent free mild cleaning agent 10 Warranty and Repairs The calibrator micro calibration bath is under...

Страница 27: ...of used metal block calibrators micro calibration baths Allow the instrument to cool down as described in chapter 8 Switch off the calibrator micro calibration bath and disconnect the mains plug If ne...

Страница 28: ...ion 0 1 C Ambient temp influence 0 50 C to the accuracy 0 02 C C Excess temperature behaviour temperature fuses interrupt the power supply if there is excess temperature inside the housing Sensor brea...

Страница 29: ...RS485 optoisolated Communication protocol MODBUS RTU JBUS Serial transfer speed 9600 baud Test specimen holder CTB9100 165 CTB9100 225 bore 60 mm depth 150 mm Housing dimensions CTB9100 165 210 mm x 4...

Страница 30: ...temp influence 0 50 C to the accuracy 0 02 C C Excess temperature behaviour temperature fuses interrupt the power supply if there is excess temperature inside the housing Sensor break behaviour the co...

Страница 31: ...al interface type RS485 optoisolated Communication protocol MODBUS RTU JBUS Serial transfer speed 9600 baud Test specimen holder CTD9100 165 bore 28 mm depth 150 mm CTD9100 450 bore 60 mm depth 150 mm...

Страница 32: ...for thermometers to 6 3 mm diameter 7 1x 7 5 mm for thermometesr to 7 2 mm diameter 8 1x 8 5 mm for thermometers to 8 2 mm diameter 9 1x 10 mm for thermometers to 9 5 mm diameter A 1 x 3 2 mm and 1 x...

Страница 33: ...min 0 C to 15 C 10 36 min 75 C to 100 C 4 23 min 15 C to 25 C 15 01 min 100 C to 120 C 5 05 min 25 C to 30 C 23 19 min GB Temperature Dry Well Calibrators Micro Calibration Baths CTD9100 CTB9100 1126...

Страница 34: ...C 2 59 min 75 C to 50 C 4 33 min 50 C to 75 C 4 17 min 50 C to 25 C 5 57 min 75 C to 100 C 5 18 min 25 C to 0 C 7 49 min 100 C to 125 C 7 09 min 0 C to 15 C 10 17 min 125 C to 150 C 12 06 min 15 C to...

Страница 35: ...22 min 150 C to 125 C 3 17 min 50 C to 75 C 0 52 min 125 C to 100 C 4 14 min 75 C to 100 C 0 53 min 100 C to 75 C 5 59 min 100 C to 125 C 0 59 min 75 C to 50 C 9 59 min 125 C to 150 C 1 12 min 50 C t...

Страница 36: ...02 min Silicone oil 50 CS Heating up Period Cooling down Period 25 C to 40 C 1 18 min 225 C to 200 C 2 37 min 40 C to 50 C 0 21 min 200 C to 165 C 3 25 min 50 C to 75 C 0 48 min 165 C to 150 C 1 47 mi...

Страница 37: ...3 min 150 C to 100 C 18 00 min 300 C to 350 C 1 33 min 100 C to 50 C 37 01 min 350 C to 400 C 1 53 min 50 C to 40 C 15 45 min 400 C to 450 C 2 33 min 40 C to 25 C 50 53 min GB Temperature Dry Well Cal...

Страница 38: ...50 C to 500 C 1 32 min 150 C to 100 C 17 01 min 500 C to 550 C 1 38 min 100 C to 50 C 52 37 min 550 C to 600 C 1 55 min 50 C to 40 C 15 23 min 600 C to 650 C 2 33 min 40 C to 25 C 1 01 58 min GB Tempe...

Страница 39: ...39 GB Temperature Dry Well Calibrators Micro Calibration Baths CTD9100 CTB9100 11263911 04 2007 GB D WIKA Operating Instructions Temperature Dry Well Calibrators Micro Calibration Baths...

Страница 40: ...PL Instrukcja obs ugi WIKA 40 Informacje Symbol wskazuje informacje uwagi i wskaz wki Ostro nie G wny punk zagro enia Nale y post powa zgodnie z instrukcj obs ugi Niebezpiecze stwo Niebezpiecze stwo u...

Страница 41: ...wanie i sprawdzenie urz dzenia po dostawie 46 4 Opis kontrolek 47 5 Uruchamianie kalibratora kalibratora k pielowego mikro 49 6 Obs uga kalibratora kalibratora k pielowego mikro 52 7 Komunikaty b du 6...

Страница 42: ...iniejsza instrukcja obs ugi dotyczy poni szych typ w kalibrator w 1 Opis urz dzenia i przeznaczenie Kalibrator temperatury CTD9100 165 ch odzenie i grzanie CTD9100 450 grzanie CTD9100 650 grzanie Kali...

Страница 43: ...przodu Ty Z ty u urz dzenia znajduje si nalepka typu zawieraj ca wa ne informacje o poszczeg lnych modelach CTD9100 165 zakres 30 C 165 C CTD9100 450 zakres 40 C 450 C CTD9100 650 zakres 40 C 650 C Wy...

Страница 44: ...ne pr dem sieciowym o niebezpiecznym napi ciu Nieprawid owe u ycie mo e spowodowa uszkodzenia cia a Do zapewnienia prawid owej i bezpiecznej pracy kalibratora jest prawid owy transport przechowywanie...

Страница 45: ...osowa pokryw transportow do zamkni cia wanny kalibracyjnej po u yciu Pokrywa transportowa jest wyposa ona w zaw r bezpiecze stwa Je eli zostanie zamkni ta jeszcze ciep a mikro wanna kalibracyjna mo e...

Страница 46: ...hronnym Nale y zachowa opakowanie ochronne do wysy ki przyrz du w celu rekalibracji lub naprawy do producenta Najpierw nale y sprawdzi stan dostawy Standardowa dostawa kalibratora temperatury Kalibrat...

Страница 47: ...a Element grzewczy Sygnalizuje status wyj cia sterowania wentylatora Je eli dioda LED OUT 2 jest zapalona wentylator pracuje z du pr dko ci Je eli dioda LED OUT 2 nie jest zapalona wentylator pracuje...

Страница 48: ...na sygna y RS 232 lub USB Odpowiednie konwertery razem ze sterownikami dost pne s jako opcje Komputer osobisty PC rejestruje dane robocze i umo liwia programowanie ca ej konfigu racji parametr w kalib...

Страница 49: ...tulejek w bloku grzewczym 5 3 Przygotowanie kalibratora k pielowego mikro W celu uzyskania najwy szej mo liwie dok adno ci mikro wanny kalibracyjnej nale y nape ni j odpow iednim p ynem kalibracyjnym...

Страница 50: ...re mo e uszkodzi wann z p ynem Gdy ci nienie osi gnie oko o 2 5 bar zostaje uruchomiony zaw r bezpiecze stwa Dzi ki temu mo na zapobiec uwalnianiu si gor cej pary 5 3 3 Bedienung des Magnetr hrers Di...

Страница 51: ...era na g rnym wy wietlaczu PV pojawi si komunikat tESt Na dolnym wy wietlaczu SV pojawi si numer wersji np rL 2 2 Inicjalizacja zostaje zako czona po oko o 5 sekundach a nast pnie automatycznie zostaj...

Страница 52: ...6 Wy wietlony tryb kalibracji wentylatora FAN lub ch odzenia COOLING a Podgrzewanie przyrz du Dioda LED OUT 2 sygnalizuje status wyj cia stero wania wentylatora Je eli dioda LED OUT 2 jest zapalona we...

Страница 53: ...sekund warto szybko ro nie lub maleje a po przytrzymaniu przez dwie sekundy jeszcze szybciej dzi ki czemu szybciej mo na osi gn wybran warto Je eli przez oko o 15 sekund nie b dzie przyci ni ty aden p...

Страница 54: ...7 GB PL Instrukcja obs ugi WIKA 54 Tak jak jest wy wietlone w strukturze menu mo na osi gn poziom grupy lub parametru w celu ustawienia poprzez OPEr Powr t do poprzedniego poziomu Je eli przez oko o 1...

Страница 55: ...jawia si na przemian wy wietlacz z aktualn temperatur wzorcow oraz z komunikatem OFF na g rnym wy wietlaczu PV Aktualna nastawa temperatury pojawia si na dolnym wy wietlaczu SV Sterowanie jest wy czon...

Страница 56: ...si na aktualna temperatura wzorcowa Na dolnym wy wietlaczu SV pojawia si litera H i aktualna nastawa wydajno ci wyj ciowej w Nacisn przycisk aby zwi kszy wydajno wyj ciow Nacisn przycisk aby zmniejszy...

Страница 57: ...j pami warto ci nastaw SP1 oraz LED SET Zastosowa przycisk lub aby wybra jedn z czterech pami ci warto ci nastaw SP1 SP2 SP3 i SP4 Nacisn ponownie przycisk P aby otworzy odpowiedni pami warto ci nasta...

Страница 58: ...by pobra inn zapami tan warto nastawy nacisn ponownie przycisk U Wybrana warto temperatury zostaje natychmiast przyj ta i zasto sowana 6 3 7 Ustawianie kontroli gradientu oraz profilu temperatury Mo n...

Страница 59: ...ona temperatura SP1 Nast pnie kalibrator mikro wanna kalibracyjna automatycznie prze cza si na nastaw temperatury SP2 Po wprowadzeniu ustawie dla podanych trzech nastaw kalibrator mikro wanna kalibrac...

Страница 60: ...a gradientu ogrzewania Nacisn przycisk aby zwi kszy gradient ogrzewania SLor Nacisn przycisk aby zmniejszy gradient ogrzewania SLor Nacisn przycisk i aby odpowiednio zwi kszy lub zmniejszy warto o 0 1...

Страница 61: ...i przyciski zostan przyci ni te i przytrzymane przez przynajmniej jedn sekund warto szybko ro nie lub maleje a po przytrzymaniu przez dwie sekundy jeszcze szybciej dzi ki czemu szybciej mo na osi gn w...

Страница 62: ...przycisk aby zmniejszy czas trwania dur t Nacisn przycisk i aby odpowiednio zwi kszy lub zmniejszy warto o 0 1 Je eli przyciski zostan przyci ni te i przytrzymane przez przynajmniej jedn sekund warto...

Страница 63: ...kcja obs ugi WIKA 7 Komunikaty b du Je eli jest konieczna naprawa nale y zamkn kalibrator mikro wann kalibracyjn patrz rozdz 12 i zwr ci urz dzenie do producenta 8 Sch adzanie bloku metalowego wanny z...

Страница 64: ...pozostawi do ca kowitego osuszenia klatk czujnika mieszad o magnetyczne i zbiornik Czyszczenie z zewn trz Oczy ci z zewn trz kalibrator mikro wann kalibracyjn wilgotn szmatk z wod lub agodnym rodkiem...

Страница 65: ...alibrator w i mikro wanien kalibracyjnych Pozostawi urz dzenie do och odzenia tak jak podano w rozdz 8 Wy czy kalibrator mikro wann kalibracyjn i wyj wtyczk zasilania elektrycznego z gniazdka Je eli j...

Страница 66: ...atury pracy 0 50 C na dok adno 0 02 C C Zachowanie w razie nadmiernej temperatury bezpieczniki temperatury przerywaj zasilanie elektryczne przy wyst pieniu nadmiernej temperatury wewn trz obudowy Zach...

Страница 67: ...485 ze sprz gaczem optycznym Protok komunikacji MODBUS RTU JBUS Pr dko transmisji szeregowej 9600 baud w Uchwyt pr by testowej CTB9100 165 CTB9100 225 otw r 60 mm g boko 150 mm Wymiary obudowy CTB9100...

Страница 68: ...no ci 0 02 C C Zachowanie w razie nadmiernej temperatury bezpieczniki temperatury przerywaj zasilanie elektryczne przy wyst pieniu nadmiernej temperatury wewn trz obudowy Zachowanie przy uszkodzeniu c...

Страница 69: ...20 Typ interfejsu szeregowego RS485 ze sprz gaczem optycznym Protok komunikacji MODBUS RTU JBUS Pr dko transmisji szerogowej 9600 baud w Uchwyt pr by testowej TP165S otw r 28 mm g boko 150 mm TP450S o...

Страница 70: ...do 4 7 mm 6 1x 6 5 mm dla termometr w o rednicy do 6 3 mm 7 1x 7 5 mm dla termometr w o rednicy do 7 2 mm 8 1x 8 5 mm dla termometr w o rednicy do 8 2 mm 9 1x 10 mm dla termometr w o rednicy do 9 5 m...

Страница 71: ...0 min 50 C do 75 C 3 10 min 0 C do 15 C 10 36 min 75 C do 100 C 4 23 min 15 C do 25 C 15 01 min 100 C do 120 C 5 05 min 25 C do 30 C 23 19 min PL 71 Kalibrator temperatury z suchym otworem pomiarowym...

Страница 72: ...47 min 25 C do 50 C 2 59 min 75 C do 50 C 4 33 min 50 C do 75 C 4 17 min 50 C do 25 C 5 57 min 75 C do 100 C 5 18 min 25 C do 0 C 7 49 min 100 C do 125 C 7 09 min 0 C do 15 C 10 17 min 125 C do 150 C...

Страница 73: ...50 C 1 40 min 40 C do 50 C 0 22 min 150 C do 125 C 3 17 min 50 C do 75 C 0 52 min 125 C do 100 C 4 14 min 75 C do 100 C 0 53 min 100 C do 75 C 5 59 min 100 C do 125 C 0 59 min 75 C do 50 C 9 59 min 12...

Страница 74: ...min 40 C do 25 C 40 02 min Olej silikonowy 50 CS Podgrzewanie Czas Sch adzanie Czas 25 C do 40 C 1 18 min 225 C do 200 C 2 37 min 40 C do 50 C 0 21 min 200 C do 165 C 3 25 min 50 C do 75 C 0 48 min 16...

Страница 75: ...12 07 min 250 C do 300 C 1 23 min 150 C do 100 C 18 00 min 300 C do 350 C 1 33 min 100 C do 50 C 37 01 min 350 C do 400 C 1 53 min 50 C do 40 C 15 45 min 400 C do 450 C 2 33 min 40 C do 25 C 50 53 min...

Страница 76: ...200 C do 150 C 11 22 min 450 C do 500 C 1 32 min 150 C do 100 C 17 01 min 500 C do 550 C 1 38 min 100 C do 50 C 52 37 min 550 C do 600 C 1 55 min 50 C do 40 C 15 23 min 600 C do 650 C 2 33 min 40 C do...

Страница 77: ...PL 77 Kalibrator temperatury z suchym otworem pomiarowym Kalibrator k pielowy mikro CTD9100 CTB9100 11263911 04 2007 GB PL Instrukcja obs ugi WIKA Notatki...

Страница 78: ...PL Kalibrator temperatury z suchym otworem pomiarowym Kalibrator k pielowy mikro CTD9100 CTB9100 11263911 04 2007 GB PL Instrukcja obs ugi WIKA 78 Notatki...

Страница 79: ...PL 79 Kalibrator temperatury z suchym otworem pomiarowym Kalibrator k pielowy mikro CTD9100 CTB9100 11263911 04 2007 GB PL Instrukcja obs ugi WIKA Notatki...

Страница 80: ...KA Alexander Wiegand SE Co KG Alexander Wiegand Stra e 30 63911 Klingenberg Germany Tel 49 9372 132 0 Fax 49 9372 132 406 E Mail info wika de www wika de Oddzia y WIKA na wiecie dost pne s na stronie...

Отзывы: