1
Calibrator type
(heating/cooling)
Setting for
“SLoF”
CTD9100-165
< 5 °C/min
CTB9100-165
with silicone oil 20 CS
< 6 °C/min
CTB9100-165
with distilLED water
< 4 °C/min
Calibrator type
(heating)
Setting for
“SLoF”
CTD9100-450, -650S
up to 300 °C
300 °C up to 100 °C
< 10 °C/min
< 5 °C/min
CTB9100-225
with silicone oil 20 CS
200 °C up to 50 °C
50 °C up to 30 °C
< 4 °C/min
< 0.5 °C/min
CTB9100-225
with distilLED water
90 °C up to 50 °C
50 °C up to 30 °C
< 2 °C/min
< 0.5 °C/min
GB
Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
11263911 04/2007 GB/D
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
If you have carried out settings for
these three settings, the calibrator/
micro calibration bath uses the new
values only when modifying the
set temperature or switching the
calibrator off and on again.
A further procedure is to switch
off the automatic control prior to
modifying parameters (see section
6..1), and to switch it on again
afterwards (see section 6..).
The heating and cooling gradients and the duration
time can be set in the parameter level ’rEG.
This is achieved by
Pressing the
P
key for approx. 5 sec., the main
menu opens.
OPEr
appears on the upper
PV
display.
LED SET flashes on the lower
SV
display.
Press the
P
key again, the
group level
opens.
OPEr
appears on the upper
PV
display.
’SP
appears on the lower
SV
display and LED
SET flashes.
Press the
P
key again, the
parameter level
opens.
’rEG
appears on the upper
PV
display.
SLor
flashes on the lower
SV
display.
Fig. 6: Operator menu OPEr
Fig. 7: Group ’SP
Fig. 8: Group ’rEG
Fig. 9: Parameters for heating gradient SLor
Use the
key to select the group
’rEG
.
OPEr
appears on the upper
PV
display.
’rEG
appears on the lower
SV
display and LED
SET flashes.