5
DistilLED water
Heating up:
Period
Cooling down:
Period
25 °C to 40 °C
0:55 min
90 °C to 75 °C
5:53 min
40 °C to 50 °C
0:37 min
75 °C to 50 °C
15:17 min
50 °C to 75 °C
1:27 min
50 °C to 40 °C
10:50 min
75 °C to 90 °C
1:30 min
40 °C to 25 °C
45:26 min
Silicone oil 5 CS
Heating up:
Period
Cooling down:
Period
25 °C to 40 °C
0:51 min
120 °C to 100 °C
3:27 min
40 °C to 50 °C
0:16 min
100 °C to 75 °C
5:55 min
50 °C to 75 °C
0:54 min
75 °C to 50 °C
10:00 min
75 °C to 100 °C
1:13 min
50 °C to 40 °C
7:02 min
100 °C to 120 °C
1:35 min
40 °C to 25 °C
34:28 min
Silicone oil 10 CS
Heating up:
Period
Cooling down:
Period
25 °C to 40 °C
0:52 min
165 °C to 150 °C
1:40 min
40 °C to 50 °C
0:22 min
150 °C to 125 °C
3:17 min
50 °C to 75 °C
0:52 min
125 °C to 100 °C
4:14 min
75 °C to 100 °C
0:53 min
100 °C to 75 °C
5:59 min
100 °C to 125 °C
0:59 min
75 °C to 50 °C
9:59 min
125 °C to 150 °C
1:12 min
50 °C to 40 °C
7:00 min
150 °C to 165 °C
1:03 min
40 °C to 25 °C
31:40 min
GB
Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
11263911 04/2007 GB/D
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
1.4. CTB9100-5
Measuring conditions:
·All heating and cooling periods are related to a
reference Pt100 probe, Ø 6 mm, stainless steel
sheath
·The reference sensor is located 5 mm above the
protective basket in the middle of the tank
·All heating and cooling periods do not include
stabilisation time
·The measurements were carried out at a room
temperature of approx. 23 °C with bath without
cover