CTD9100-650
Heating up:
Period
Cooling down:
Period
25 °C to 40 °C
0:54 min
650 °C to 600 °C
2:25 min
40 °C to 50 °C
0:22 min
600 °C to 550 °C
2:33 min
50 °C to 100 °C
1:18 min
550 °C to 500 °C
2:55 min
100 °C to 150 °C
1:06 min
500 °C to 450 °C
3:27 min
150 °C to 200 °C
1:03 min
450 °C to 400 °C
4:01 min
200 °C to 250 °C
1:05 min
400 °C to 350 °C
4:39 min
250 °C to 300 °C
1:06 min
350 °C to 300 °C
5:36 min
300 °C to 350 °C
1:09 min
300 °C to 250 °C
6:46 min
350 °C to 400 °C
1:21 min
250 °C to 200 °C
8:32 min
400 °C to 450 °C
1:30 min
200 °C to 150 °C
11:22 min
450 °C to 500 °C
1:32 min
150 °C to 100 °C
17:01 min
500 °C to 550 °C
1:38 min
100 °C to 50 °C
52:37 min
550 °C to 600 °C
1:55 min
50 °C to 40 °C
15:23 min
600 °C to 650 °C
2:33 min
40 °C to 25 °C
1:01:58 min
GB
Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
11263911 04/2007 GB/D
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
8
1.4.5 CTD9100-650
Measuring conditions:
All heating and cooling periods are related to a
reference Pt100 probe, Ø 6 mm, stainless steel
sheath.
All heating and cooling periods do not include
stabilisation time
At full depth the reference sensor is located in
the middle of the sleeve
The measurements were carried out at a room
temperature of approx. 23 °C