6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET, ET-P, ET/IS, ET-P/IS 71
GB
Pressing POINTS, the list of the calibration points will be shown:
Figure 81
– Display test points Report
Pressing RESULT, the calibration results are shown:
Figure 82
– Display Report results
The buttons on the bottom of the screens allow to access further points (if present):
“As found”, “As
left”, “Up”, “Down”, will appear to enable to see the calibration results related to the different
calibration stages.
“As found”, (calibration points before any adjustment on the DUT) “As left” (calibration points after
the adjustment on the DUT has been carried out)
,“Up” (calibration points going up), “Down”
(calibration points going down).
On the screen for the Functions Setting (see paragraph 6.5.2.) press DELETE ALL to delete all
the reports in the memory.
Note
: a confirmation message will appear before deleting all data
6.6.1 Report of Tc/Rtd with dry well calibrator or dry block management
The Pascal calibrator allows to create reports for Thermocouples/Thermoresistances by using a
dry well calibrator and by automatically manage the temperature points. With the serial interface
the Pascal calibrator is able to connect to some dry well calibrators and automatically set the
temperature value e and detect the reached stability in order to automatically store the temperature
points and then run the test autonomously.
Scrolling menu
Содержание CPH8000
Страница 11: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 11 GB 3 Specifications 1...
Страница 12: ...3 Specifications 12 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...
Страница 13: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 13 GB...
Страница 14: ...3 Specifications 14 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...