6 Commissioning, operation
WIKA Operating Instruction, Model Pascal ET, ET-P, ET/IS, ET-P/IS 25
GB
6.2.1 Electrical measurements
The instrument is able to measure voltage, current, resistance and frequency. To optimize the
resolution and get better measurement results, there are three ranges for voltage measurement,
two for resistance measurements, one range for current and three ranges for frequency.
Ranges of measurement and resolutions are specified in following table:
Range
Max resolution
Voltage
(-
100…100) mV
0.0001 mV
(-2
…2) V
0.000001 V
(-
80…80) V
0.00001 V
Current
(-10
0…100) mA
0.0001 mA
Resistance
(0…400) Ω
0.001
Ω
(0…10000) Ω
0.01
Ω
Frequency
(0.
4…50000) Hz
0.001 Hz
Pulse
0,4…20 Hz
-
Contact
Open - Closed
-
Table 1 - Range of input signals