6 Commissioning, operation
48 WIKA Operating Instruction, Pascal ET, ET-P, ET/IS, ET-P/IS
GB
The calibrator is now set for the module CALC and it preset the last used calculation type for this
module (in this case the last used calculation type was ChA+ChB).
Press the button indicated with “R” in the picture above to select another calculation type.
Figure 44
– Types of calculation
Here you can find:
CHA + CHB
channel 4 or C displays the total value of the values displayed on channel A and B
CHA - CHB
channel 4 displays the difference value of the values displayed on channel A and B
Cell Load
channel 4 or D4 displays the ratio mV/V of the load cells
CHA+CHB and CHA-CHB functions can be used only if channel A and B are set with the same
values: same engineering unit, same number of digits, no channel assigned with error indication.
Otherwise the channel 4 will be displayed as:
Figure 45 - Calculation non possible for the Channel 4