6 Commissioning, operation
24 WIKA Operating Instruction, Pascal ET, ET-P, ET/IS, ET-P/IS
GB
Figure 7 -
Case 4: 24V external, 250Ω external
Figure 8 - Case 8
: 24V from module, 250Ω from module, mA reading by IN module
6.1.2.5 Environmental parameters module (optional)
This module allows to measure: temperature, relative humidity and atmospheric pressure. These
parameters represent the most important factors that affect the calibration results. This module is
connected to Pascal ET, ET-P, ET/IS, ET-P/IS with a multiple pin connector present in Pascal ET,
ET-P, ET/IS, ET-P/IS front panel. This module is plug & play and the relevant information about the
environmental parameters are displayed on the status bar at the top of the screen.
This information is automatically reported on the calibration report.
Parameter
Range
Precision
Uncertainty Max. resolution
Temperature
-10
… +50 °C
2.7 °C
3.0 °C
0,1 °C
Barometric pressure
650
… 1150) mbar
4 % fs
5 % fs
1 mbar
Relative humidity
10
… 90 %
12 %
15 %
1 %
6.2
Electrical
The Pascal ET, ET-P, ET/IS, ET-P/IS are multifunction calibrators.
The electrical modules can be maximum four, two of them are input modules (IN A
– IN B), and the
other two are output modules (OUT A
– OUT B).
Transmitter
HART slave
24V ext.
250
Ω
Transmitter
HART slave
Содержание CPH8000
Страница 11: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 11 GB 3 Specifications 1...
Страница 12: ...3 Specifications 12 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...
Страница 13: ...3 Specifications WIKA Operating Instruction Model Pascal ET ET P ET IS ET P IS 13 GB...
Страница 14: ...3 Specifications 14 WIKA Operating Instruction Pascal ET ET P ET IS ET P IS GB...