6 Commissioning, operation
46 WIKA Operating Instruction, Pascal ET, ET-P, ET/IS, ET-P/IS
GB
6.3.2.4 Temperature Simulation
The procedure for the temperature simulation assignment is similar to the measurement one as
described previously, except for the step where the operator needs to select operative module by
pressing the button indicated with “Q”:
Figure 39
– Selecting channel assignment OUT A - mA
The next step is the selection of one of the output modules OUT A or OUT B (if present):
Figure 40
– Selecting generation module (OUT A or OUT B)
6.3.2.5 Electrical parameter generation
The procedure for the assignment of the generation of an electric parameter (for example mA) is
similar to the electrical signal measurement one as described previously, expect for the step where
the operator needs to select operative module by pressing the button indicated with “Q”.
In this case the option OUT A or OUT B should be selected.
From the main screen, now the loop type must be selected between: passive (the calibrator Pascal
supply the current loop) or active (current loop supplied externally):
Q