Pump head:

 

Other optional pumpheads can be mounted on the unit, (see section 6.1). Tubing is loaded 

by simply raising the cover on the pumphead, and snapping closed 

 

5.1  Control 

The  OmniSpense

®

 ELITE can be used in both laboratory and industrial settings. In addition to the front-

panel keypad, the pump can be triggered to run remotely either by the optional footswitch, or by a dry 

contact  relay  or  isolated  open  collector  transistor  output  as  part  of  a  more  advanced  automated 

dispensing system. The  OmniSpense

®

 ELITE is capable of sending an End of Dispense (EOD) feedback 

signal to a controller for remote automated operation. 
 

5.2  User Interface 

A chemical resistant membrane keypad allows direct user entry of volume, tubing size, automatic delay 

interval, ramp time, pump direction, pump speed, pump function and dose count. Additional user selective 

functions include calibration, memory functions and audible cycle alert. The self-prompting, symbolic LCD 

screen ensures quick setup and ease of operation.  
 

5.3  Performance Highlights 

An adjustable ramp feature allows the pump to come up to speed gradually and helps prevent foaming or 

splashing  when dispensing into dishes  or small vials. An adjustable delay can  be set from 0.5 to  99.9 

seconds between doses to accommodate user dexterity and work load. The pump can be activated by 

depressing either the front panel mounted start key, or the optional footswitch for hands-free operation. 
Up to 9 pre-selected setups can be stored in memory and quickly retrieved for later use. Selections stored 

in memory are preserved even after powering the unit down. Pump speed is variable from .5 to 400 RPM, 

which facilitates usage where flow rate control is critical.  A flexible tubing support stand is included to 

facilitate hands-free operation. 

 

6.0  Installation 

 

Install  the  unit  where  there  will  be  adequate  room  for  the  unit  to  operate.    Provide  enough  clearance 

around the unit to keep items away from the rotating pumps and tubing.  

 

6.1  Input Power Requirements 

This equipment is designed to operate from a 100-120V single-phase AC power source at 47 to 63 Hz.  

The line voltage / fuse label located on the lower rear of the unit shows the nominal input voltage set for 

the unit at the factory.    
 

6.2  Power Cord Set 

This unit has been shipped from the factory with a power line cord that has a plug appropriate for your 

area.  If  the  wrong  power  cord  has  been  shipped  for  your  particular  application,  contact  your  nearest 

Wheaton Industries dealer for the proper cord. The  OmniSpense

®

 ELITE has been equipped with a 3-

wire  grounding  type  power  cord.  The  unit  is  only  grounded  when  it  is  plugged  into  an  appropriate 

receptacle. 

Do not operate the unit without adequate grounding protection. 

 

 

Содержание OmniSpense ELITE Series

Страница 1: ...WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 Catalog Numbers W375030 A and W375030 B 100 120V And W375030 C through W375030 J 220 240V...

Страница 2: ...terface 5 3 Performance Highlights 6 0 Installation 6 1 Input Power Requirements 6 2 Power Cord Set 7 0 Initial Startup and Operation 7 1 Pump Head Attachment and Tubing Installation 7 2 Tubing Charac...

Страница 3: ...gs 12 0 Constant Flow Rate Dispensing 12 1 Constant Flow Rate 12 2 Soft Start Ramping 12 3 Changing Flow Rate On The Fly 13 0 Working with Alternate Pump Head Setups 13 1 Multiple Pump Heads 13 2 Work...

Страница 4: ...of identifying such a problem by calling Customer Service at 800 225 1437 U S Puerto Rico and Canada or Internationally at 856 825 1100 and then providing the Model Serial and Catalog Number of the p...

Страница 5: ...t caught in moving parts Non slip footwear is recommended Wear protective hair covering to contain long hair 6 Wear safety goggles Wear safety goggles at all times Everyday eyeglasses only have impact...

Страница 6: ...pumphead 7 H x 8 W x 13 D in 17 8H x 22 3W x 33 0D cm Weight 11 8 lbs 5 4 kg Available Wheaton Tubing Sizes 2 3 6 8 mm ID Maximum Dispense Range for All Tubing Sizes 01 9999 99 ml Approximate Flow Rat...

Страница 7: ...Initial Inspection When you receive your new OmniSpense ELITE inspect it for any obvious damage that may have occurred during shipment If any damage is found notify the carrier at once Warranty inform...

Страница 8: ...adjustable delay can be set from 0 5 to 99 9 seconds between doses to accommodate user dexterity and work load The pump can be activated by depressing either the front panel mounted start key or the o...

Страница 9: ...ting plate Place the pump head on the mounting plate and twist the pump head clockwise until it clicks into place Tubing Installation Insert Tubing into the pump head between the pump head cover and t...

Страница 10: ...3 Switching the Unit ON Switch the pump ON via the power switch on the rear panel The main screen displays the Wheaton Industries logo and the firmware version level The pump will automatically trans...

Страница 11: ...a Pump Running Pump Stopped indicator Pins 1 footswitch and 2 footswitch GND are 5V TTL compatible inputs with pull up resistors and can be driven by an isolated open collector device or dry contact...

Страница 12: ...ON will cause the unit to sound an end of dispense alert Turning the buzzer function OFF shown will disable the buzzer during the dispense cycle 8 3 Unit Reset Press the UP and DOWN and the LEFT and R...

Страница 13: ...Cassette Pump Model 304 308 9 2 Pump Direction Use the UP and DOWN keys to move to the pump direction icon Use the LEFT and RIGHT arrow keys to select either clockwise or counterclockwise direction C...

Страница 14: ...n The PRIME button on the keypad now becomes active Push and HOLD the prime button on the keypad to run the pump motor and purge any air from the tubing RELEASE the prime button to stop the motor 10 3...

Страница 15: ...olume Press OK to enter the volume and complete the calibration procedure Press OK again to return to the HOME screen 11 0 Making Batch and Single Dose Dispenses From the HOME screen press OK to navig...

Страница 16: ...RIGHT keys to enter volume per dispense Press OK when done 11 4 Delay Between Dispenses Use the UP and DOWN keys to navigate to the Delay icon Use the UP and DOWN and the LEFT and RIGHT keys to enter...

Страница 17: ...k Volume set to 5 mL Use the UP and DOWN and the LEFT and RIGHT keys to enter the drawback volume Press OK when done 11 6 Soft Start Ramping Use the UP and DOWN keys to navigate to the Ramp icon Press...

Страница 18: ...uring a batch dispense by pressing the Start Stop button and then using the UP and DOWN and the LEFT and RIGHT keys to enter the dispense volume Press OK when done Press the Start Stop button to resum...

Страница 19: ...800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 19 Flow Mode Press OK to navigate to the FLOW SET UP screen Set Flow Rate Maximum attainable flow rate base...

Страница 20: ...The Fly The flow rate can be changed while the unit is running by using the LEFT and RIGHT keys to navigate to the running flow rate then using the UP and DOWN and the LEFT and RIGHT keys to enter th...

Страница 21: ...th Y couplers into a single outlet The combined volume of the two pump heads is used in the calibration number Running two pump heads as two independent processes is not recommended as a single calibr...

Страница 22: ...quickly recalled and the pump made ready for a particular operation 14 1 To Store a Parameter Set all pump parameters to desired settings Navigate to the HOME screen and select the storage icon Use th...

Страница 23: ...vide a different liquid delivery range when used in conjunction with the pump unit Accuracy is increased as the number of pump head revolutions required per dispense is increased Tubing size should be...

Страница 24: ...cy ramp rates should be set to as long as can be tolerated as a long ramp rate will naturally lengthen the dispense cycle time 15 3 Using Multiple Pump Heads Successful use of multiple pump heads depe...

Страница 25: ...d warm air Although the OmniSpense ELITE is designed for easy service it is highly recommended that the instrument be returned to Wheaton Industries should any major repairs be needed Pump rollers can...

Страница 26: ...mbly Motor fails to maintain constant rotation speed Cause Motor control board defective Remedy Have defective control board replaced Cause Drive motor defective Remedy Have defective motor replaced L...

Страница 27: ...unit 3 Replace with new fuses 108 120 vac unit use 2 5x20mm 250V 2 0 AT fuses be sure the correct cord set is used 220 240 vac unit use 2 5x20 mm 240V 1 0 AT fuses be sure the correct cord set is use...

Страница 28: ...istaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 28 FIGURE A Major Component Identification Display Keypad Pump Head Remote I O Connect...

Страница 29: ...1 50029919 NUT HEX 6 32 16 1 I052273 WASHER STAR 6 INT TOOTH 17 1 50099063 STF 6 32 X 1 4 ALUMINUM 18 2 WI056864 FUSE 5X20MM 2 0AT 250V 120V 18 2 WI056862 FUSE 5X20MM 1 0AT 250V FIGURES C THRU E Uppe...

Страница 30: ...WHEATON OmniSpense ELITE Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 30...

Страница 31: ...WHEATON OmniSpense ELITE Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 31...

Страница 32: ...WHEATON OmniSpense ELITE Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 32...

Страница 33: ...WHEATON OmniSpense ELITE Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 33...

Страница 34: ...WHEATON OmniSpense ELITE Peristaltic Pump WHEATON 800 225 1437 U S Canada Only 856 825 1100 www wheaton com P N 50099067 Rev 5 15 12 WMAN_001 34...

Страница 35: ...device described below marked with CE fulfills the relevant fundamental EMC and safety requirements specified by the appropriate EU Directive with respect to the design and construction of the commer...

Отзывы: